    Dengan hormat,

    Bersama ini, kami PT Fajar Mas Murni sebagai agen tunggal Olympus Microscope, Hitachi ScanningElectron microscope, Sakura Finetek Histopathology Equipment, Articko Freezer & Refrigerator, danFASTER Safety Cabinet di Indonesia dengan senang hati menyampaikan informasi harga dan spesifikasiteknis seperti terlampir.

    Kami harapkan referensi harga ini dapat memenuhi kebutuhan di instansi Ibu/Bapak, apabila dibutuhkan

    keterangan lebih lanjut kami mohon jangan segan-segan untuk menghubungi kami.

    Demikian referensi harga kami, sambil menanti informasi lebih lanjut dari Ibu/Bapak, kami ucapkan banyakterima kasih.

    Hormat kami,PT FAJAR MAS MURNI

    Anton Sukoco, STKepala Cabang

    note : harga berlaku apabila maksimal Kurs 1 Euro =Rp. 17.000,-

    Contact person : Purwanto

    cell phone :+62-81331876223

    email :[email protected]

  • 8/19/2019 Informasi Harga CytoFast Elite 212

    2/3

     

    Address : Jl.Jemur Andayani No.27 - Surabaya 60237, East Java, IndonesiaPhone : (62-31) 843 9827 (Hunting), Fax : (62-31) 849 5224,843 8405

    E-mail : [email protected], Web-Site : http://fajarmasmurni.com

    Lampiran :

    No Spesifikasi Harga Satuan

    1 Cytotoxic Safety Cabinet merk Faster model CytoFAST Elite 212 Rp 286.365.000

    Cabinet has been designed to protect both the material to be manipulated fromcontamination and to protect the operator and the environment from cytotoxiccontamination and harmful agents. Manufactured according to EN 12469:2000 andDIN 12980:2005 standards, with about 70% laminar flow recirculated air and 30%exhausted air with prior HEPA filtering. Third bank of class H14 HEPA filters witheasy replace system is positioned beneath work surface Built for handling cytotoxicdrugs CytoFAST Elite is also suitable for product, personnel and environmentprotection while handling harmful agents pathogenic to human beings and/or animalsas definedin the appropriate international standards.Applications : Preparation and handling of cytotoxic drugs, preparations of CMRdrugs, presence of agents likely to cause genetic alterations or synergetic activitieswith other materials. Cabinets meet also EN12469:2000 as for Class IIMicrobiological Safety Cabinets, they can be used also for: Microbiology, Virology,Haematology, Cell culture, Recombinant DNA.Ergonomic Design: The angled sloping front stratified safety-glass sash providesoptimum visibility of all objects placed in the interior workspace. The sash iselectrically operated. Pressing the appropriate touch-sensitive keys will completelyopen or completely close down the sash. The standard sash-height opening duringwork is set to 200 mm. Alternative sash-height settings by the factory are possibleupon request. The whole front of the safety cabinet  – which includes the sashmechanism – can also be opened upwards as it is hinged on the top  – in order toallow easy access for complete and effective cleaning and decontamination.

    ECS® Eco Controling System: The new ECS® microprocessor employs the latestinnovative methods of integrated management of all principal functions of ventilationand filtration - self-regulating all the main filtration and ventilation systemcomponents - compensating for declining pressure drops and restoring powerbalance. Combining the use of AC motor-blowers and certified low pressure-dropfilters, the new ECS® controlling system optimize power consumption, reducing CO2emissions into the environmentAnti Bacterial Coating: Exclusive Dupont™ ALESTA® anti-bacterial “Ag+ cations-based solution”, capable to prevent microbial contamination of surfaces therebyinhibiting long term surface growthInflatable gasketing is also available as an option to provide effective sealing of theinternal workspace of the chamber for purposes of fumigation/sterilization and whengas-tight is needed

    Silent Operation. The TNT plenum, the structures of the electric motors of the fansfitted on their antivibration mounts and the software itself designed to provideoptimum air handling characteristics guarantee quiet operation of this silent safetycabinet, with sound-pressure levels recorded way below the parameters specified inthe current EN:12469 European Standard for Microbiological Safety Cabinets.Protection grid. A protection grid made of AISI 304L stainless steel is providedinside the part behind the rear panel (where air flows), in order to avoid the tissueused during cleaning operations be sucked in the motor-blower causing the failure of

  • 8/19/2019 Informasi Harga CytoFast Elite 212

    3/3

     

    Address : Jl.Jemur Andayani No.27 - Surabaya 60237, East Java, IndonesiaPhone : (62-31) 843 9827 (Hunting), Fax : (62-31) 849 5224,843 8405

    E-mail : [email protected], Web-Site : http://fajarmasmurni.com

    the blower and other damagesHigh Level Lighting. The safety glass side-windows with the ideal positioning andsizing of the light-system provide the highest level of luminosity to the work area.Easy handling and maintenance. The safety cabinet can pass through standard800 mm wide door openings. In fact, the overall depth of the cabinet can be reducedto approx. 790 mm by removal of the rear panel. All service operations are availablefrom the front of the cabinet.Demountable base stand of 255 mm. Without this stand the cabinet height is 1995mm and can easily pass through doors of 2 meters height.Rear wall in stainless steel AISI 304 L, designed to conform to requirements andpass the “cleanability test” according to EN12469:2000 and DIN 12980:2005.   AISI 316L work surface consisting of sections which are easily removable forcarrying out routine cleaning and/or autoclaving sterilization procedures; closed orperforated on request.

    Re-circulating and extractor fans: supplied with double centrifugal fan ( main andexhaust ) to provide complete operator, product and environmental protection.Moreover are designed and are also suitable to discharge the filtered air outside thelaboratory through a ducting system if required.Operation Condition: Class ISO 3 as per ISO:EN 14644-1The user-friendly practical keyboard and the rear-lit LCD will continuously displayall required data keeping the user constantly informed of the cabinet conditions inoperation, and in particular:

      display of laminar airflow velocity and frontal air barrier velocity

      display of inside and outside temperature

      display of residual lifetime of HEPA filters, UV Lamp and activated carbon filter(if fitted)

      display of total number of hours of operation

      display of saturation level of HEPA filters.

    Audio-visual alarms provided for:

      out of range or incorrect laminar airflow velocity and frontal air barrier velocity

      uncorrect position of front sash-window

      clogging of HEPA filters

      end of life-cycle of UV lamp and saturation of activated carbon filter (if fitted)

      blockage in the exhaust duct ( if fitted )

      fan-motor malfunction

     power failure

      UV sterilizing lamp (optional) complete two switch-off countdown timers, onevariable on a 0-3 hours scale (1 minute steps), the other set to 3 fixed hours.

    Magnetic U.V. Lamp with full programable timer .

    Useful Dimensions: 1192 x 740 x 580 ( W x H x D ) mmOverall Dimensions: 1350 x 2345 x 855 ( W x H x D ) mmNoise level 1200 lux, exhaust flow rate 390 m